FACSAria Fusion specifications

The BD FACSAria Fusion is an advanced system for experienced users, achieving high sensitivity and resolution with flexibility and control. The Aria Fusion can meet a variety of needs with its range of nozzle sizes, adjustable dynamic range of sample flow rates, and selection of sort devices- including 2-way and 4-way tubes as well as 6-, 24-, 48-, 96-, and 384-well plates. The Aria Fusion is installed in a fully integrated custom Class II Type A2 biosafety cabinet for protection of personnel and products. The cabinet incorporates a built-in aerosol management system which operates independently to actively evacuate aerosols from the sort collection chamber. The dual routes of aerosol evacuation maximize protection.

Excitation lasers

488 nm, 640 nm, 405nm, 561 nm

Output power

405 nm (>80 mW), 488 nm (>50 mW), 640 nm (>100mW), 561 nm (>50 mW)

Detection parameters

14 fluorescence + 2 scatter

Sample tube

Single, auto-loading tube

Tube types

1.5 mL, 2.0-mL, 5.0mL, and 15 mL

Temperature control

4, 20, 37, or 42°C

Agitation unit

Eccentric rotation

Sorting nozzle

70, 85, 100, and 130 um